FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 3771418 · Received April 24, 2014

Report

Report Number
2124215-2014-09200
Event Type
Injury
Date Received
April 24, 2014
Date of Event
January 21, 2014
Report Date
May 1, 2014
Manufacturer
CPI - DEL CARIBE
Product Code
NIK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Additional Manufacturer Narrative · 1

NO ADDITIONAL INFORMATION IS CURRENTLY AVAILABLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THE EVENT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED FUNCTIONAL UNDERSENSING AND LOSS OF CAPTURE DUE TO A DISLODGEMENT. THE DISLODGEMENT WAS CONFIRMED VIA A CHEST X-RAY. SUBSEQUENTLY THE LEAD WAS EXPLANTED AND ANOTHER MANUFACTURER'S LV LEAD WAS IMPLANTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249053 EASYTRAK 2 IMPLANTABLE LEAD NIK CPI - DEL CARIBE 4543

Patients

Seq Age Sex Outcome Treatment
1 79 YR Hospitalization| L| R 0175| N141| 4087| 4543