AIM-ARM 130° F/PFNA BLADE
Report
- Report Number
- 9612488-2014-10124
- Event Type
- Malfunction
- Date Received
- April 24, 2014
- Report Date
- March 27, 2014
- Manufacturer
- SYNTHES BETTLACH
- Product Code
- FZX
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
ADDITIONAL NARRATIVE: (B)(6). SERVICE HISTORY REVIEW WAS CONDUCTED. THE REPORT INDICATES THAT THE DEVICE HAS NOT BEEN SERVICED DURING THE PAST 6 MONTHS. THERE IS NO INFORMATION RELEVANT TO THE CURRENT COMPLAINT ISSUE. THE INVESTIGATION IS BASED ON THE SERVICE & REPAIR RECORD RECEIVED. THE DEVICE FAILURE REPORTED BY THE CUSTOMER COULD BE CONFIRMED. THE SERVICE HISTORY REVIEW SHOWS NO PREVIOUS SERVICE CONDITIONS RELEVANT TO THE CURRENT COMPLAINT ISSUE. THE SERVICE TECHNICIAN NOTED THE FOLLOWING ACTION REPAIR. DURING THE PRE-REPAIR DIAGNOSTIC ASSESSMENT THE SERVICE TECHNICIAN IDENTIFIED THE FOLLOWING FAILURE FLAT SPRING OF THE LOCK-DEVICE IS WORN OUT. THE ROOT CAUSE OF THE EVENT WAS DETERMINED TO BE NORMAL WEAR. THE DEVICE WAS REPAIRED AND RETURNED TO CUSTOMER. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT THE ARMING BLADE DOES NOT WORK. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS REPORT IS 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 248727 | AIM-ARM 130° F/PFNA BLADE | GUIDE | FZX | SYNTHES BETTLACH | 2646885 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |