VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2014-00372
- Event Type
- Malfunction
- Date Received
- April 24, 2014
- Date of Event
- March 23, 2014
- Report Date
- March 26, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE COMPLAINT MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) WHERE IT WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED A BREAK IN THE WATER FEEDSET TUBE AT THE CONNECTION TO THE CHAMBER DOME. THE SURFACE OF THE BREAK WAS ROUGH (NOT SMOOTHLY CUT). A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120628. CONCLUSION: THE DAMAGE APPEARS TO HAVE BEEN CAUSED BY THE FEEDSET TUBE BEING PULLED AWAY FROM THE CHAMBER DOME, POSSIBLY DUE TO THE FEEDSET BEING CAUGHT OR PUT UNDER TENSION. WE HAVE CONDUCTED EXTENSIVE TESTING OF THE MR290 CHAMBER, WITH PARTICULAR EMPHASIS ON FEEDSET BREAKS. SIGNIFICANTLY WE HAVE NOT BEEN ABLE TO REPLICATE FAILURE OF THE FEEDSET TUBE AT THE CHAMBER DOME IN ANY OF OUR TESTING. THE SPECIFICATION FOR THE CHAMBER REQUIRES THAT THE FEEDSET TUBE SHOULD HAVE A BREAKING STRAIN OF 30 NEWTONS. DURING PRODUCTION, PULL TESTING OF THE FEEDSET STRENGTH AT BOTH SPIKE AND DOME END IS PERFORMED EVERY HOUR ON MR290 CHAMBERS FROM EACH PRODUCTION LINE. IF ANY PRODUCT FAILS, THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. ADDITIONALLY ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. CHAMBERS THAT FAIL ANY OF THESE TESTS ARE DISCARDED. THIS SUGGESTS THAT THE DAMAGE OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: SET APPROPRIATE VENTILATOR ALARMS; PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THE WATER FEEDSET TUBE FROM AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS DAMAGED. THIS WAS FOUND PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT THE WATER FEEDSET TUBE FROM AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS DAMAGED. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 248724 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 1206280105 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | FPH RT202 ADULT BREATHING CIRCUIT| FPH RT202 ADULT BREATHING CIRCUIT |