FDA Adverse Event Malfunction Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3771246 · Received April 24, 2014

Report

Report Number
2024168-2014-02570
Event Type
Malfunction
Date Received
April 24, 2014
Date of Event
April 10, 2014
Report Date
April 11, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE STENT DISLODGEMENT WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL AND DIMENSIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY/REVIEW OF THE ELECTRONIC COMPLAINT DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FOR STENT DISLODGEMENTS FROM THIS LOT. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING UNPACKAGING OF THE 2.75X28MM XIENCE XPEDITION STENT DELIVERY SYSTEM (SDS), THE STENT IMPLANT DISLODGED FROM THE BALLOON DURING PROTECTIVE SHEATH REMOVAL. REPORTEDLY, THERE WAS NO RESISTANCE NOTED DURING REMOVAL OF THE STYLET OR PROTECTIVE SHEATH. THERE WAS NO REPORTED PATIENT INVOLVEMENT AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ANOTHER 2.75X28MM XIENCE XPEDITION SDS WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
250253 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 210236A

Patients

Seq Age Sex Outcome Treatment
1