FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 20E DEFIBRILLATOR/MONITOR

MDR report key: 3770867 · Received April 24, 2014

Report

Report Number
3015876-2014-00461
Event Type
Malfunction
Date Received
April 24, 2014
Date of Event
March 29, 2014
Report Date
March 31, 2014
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K073089
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND CONFIRMED THE REPORTED ISSUE. THE CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE DUE TO A FAILURE OF DIODE CR30 ON THE THERAPY PCB ASSEMBLY. THE DIODE WAS SHORTED FROM PINS 5 TO 9. THE DEVICE WAS REPAIRED BY REPLACING THE THERAPY PCB ASSEMBLY. FOLLOWING THE REPAIR, PROPER OPERATION WAS CONFIRMED DURING FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS SUBSEQUENTLY RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE DEVICE WAS DISPLAYING THE SERVICE INDICATOR. UPON EVALUATION, PHYSIO-CONTROL OBSERVED THAT DEVICE HAD LOGGED EVENT CODE 9C19. THE EVENT CODE IS INDICATIVE OF A DEVICE FAILURE THAT COULD RESULT IN A PARTIAL LOSS OF DEFIBRILLATOR OUTPUT ENERGY DUE TO A LOSS OF THE NEGATIVE PORTION OF THE BIPHASIC OUTPUT WAVEFORM. THE DEFIBRILLATOR OUTPUT ENERGY COULD BE REDUCED BY UP TO APPROXIMATELY 20% FROM THE SELECTED ENERGY LEVEL. THERE WAS NO REPORT OF PATIENT USE ASSOCIATED WITH THE REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
248948 LIFEPAK(R) 20E DEFIBRILLATOR/MONITOR DEFIBRILLATORS, AUTOMATIC, EXTERNAL MKJ PHYSIO-CONTROL, INC 20E

Patients

Seq Age Sex Outcome Treatment
1