FDA Adverse Event Malfunction Summary report: N

PLM A+ SPANISH 220V

MDR report key: 3770806 · Received February 18, 2014

Report

Report Number
9615050-2014-01254
Event Type
Malfunction
Date Received
February 18, 2014
Date of Event
January 23, 2014
Report Date
January 23, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
FA212-05
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED PREVENTATIVE MAINTENANCE TESTING AT THE USER FACILITY. DURING TESTING, THE FLUID SHIELD WAS FOUND TO BE DAMAGED AND FLUID SPILLAGE WAS NOTED. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE RPTR UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE FLUID SHIELD WAS FOUND TO BE DAMAGED AND FLUID SPILLAGE WAS NOTED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
101750 PLM A+ SPANISH 220V 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA PLUM A+ SOFTWARE MODULE, LIST #UNK, SN UNK