FDA Adverse Event
Malfunction
Summary report: N
XMAX MOTOR, HIGH STRENGTH HOSE
MDR report key: 3770664
·
Received April 24, 2014
Report
- Report Number
- 1045834-2014-11875
- Event Type
- Malfunction
- Date Received
- April 24, 2014
- Date of Event
- April 9, 2014
- Report Date
- April 9, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- PK965080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
Additional Manufacturer Narrative · 1
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE NORMAL WEAR OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING ENGINEERING EVALUATION, IT WAS DISCOVERED THAT THE MOTOR DEVICE DID NOT ROTATE. THE EVENT WAS NOT RELATED TO SURGERY. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS ALLEGED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 249302 | XMAX MOTOR, HIGH STRENGTH HOSE | DRILL, SURGICAL, ENT (PNEUMATIC) - HANDPIECE | ERL | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |