FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC HEAT PACK
MDR report key: 3770280
·
Received March 7, 2014
Report
- Report Number
- 1811605-2014-00055
- Event Type
- Malfunction
- Date Received
- March 7, 2014
- Date of Event
- February 25, 2014
- Report Date
- March 6, 2014
- Manufacturer
- HEMI
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE USER REPORTED THAT THE SWITCH SPARKED. UPON FURTHER INVESTIGATION WE FOUND THAT THE CORD HAD BEEN SEVERED BY AN EXTERNAL SOURCE. OUR INVESTIGATION SHOWS THE SPARKING OCCURRED AS THE CORD WAS SEVERED. BASED ON THE OVERALL CONDITION OF THE PAD, WE CONCLUDED THAT THE HEATING PAD WAS NOT USED IN ACCORDANCE WITH OUR INSTRUCTIONS. THE PAD WAS BUNCHED UP AND DIRTY. THE LEADS WERE BENT AND BROKEN. THERMOSTATS WERE DISCOLORED.
Description of Event or Problem · 1
CONSUMER CALLED REPORTING THAT THE PAD SPARKED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 139565 | THERMOPHORE AUTOMATIC HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | HEMI | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |