SONOSURG SCISSORS 5MM O.D. HF SERIES
Report
- Report Number
- 8010047-2014-00096
- Event Type
- Malfunction
- Date Received
- March 7, 2014
- Date of Event
- February 19, 2014
- Report Date
- February 21, 2014
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- LFL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE SUBJECT DEVICE WAS RETURNED TO OMSC FOR EVAL. THE EVAL CONFIRMED THAT THE PROBE BROKE DOWN AT 15 MM FROM THE DISTAL END. THERE WERE NOT ANY SCRATCHES AT THE BROKEN POINT. THE MFG HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. AS THE RESULT OF THE EVAL, SINCE THERE WERE NOT ANY SCRATCHES AT THE BROKEN POINT, WE HAVE CONCLUDED THAT THE PROBE WAS CRACKED DUE TO ACTIVATING THE ULTRASOUND OUTPUT WITH STRONG FORCE OR TWISTING THE SUBJECT DEVICE GRASPING A TISSUE, AND EVENTUALLY BROKE DUE TO ACTIVATING THE ULTRASOUND OUTPUT CONTINUOUSLY USING THE CRACKED PROBE. IT IS CONSIDERED THAT THE ABNORMAL SOUND OCCURRED WHEN ACTIVATING THE ULTRASOUND OUTPUT WITH THE CRACKED PROBE. THE INSTRUCTION MANUAL OF SONOSURG SCISSORS MENTIONS WARNING AND CAUTION FOR POSSIBLE MISHANDLING MENTIONED ABOVE. BASED UPON THE FINDING OF EVAL, IMPROPER HANDLING THE SUBJECT DEVICE CANNOT BE RULED OUT AS A CONTRIBUTORY FACTOR TO THIS EVENT. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
OLYMPUS MEDICAL SYSTEM CORPO (OMSC) WAS INFORMED THAT DURING MUSCLE-PRESERVING MASTECTOMY, AN ABNORMAL SOUND OCCURRED AND THE PROBE OF THE SUBJECT DEVICE BROKE. THE BROKEN PROBE WAS RETRIEVED AND THE PROCEDURE WAS COMPLETED WITH OTHER DEVICE. THERE WAS NO REPORT OF INJURY REGARDING THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 139616 | SONOSURG SCISSORS 5MM O.D. HF SERIES | SONOSURG SCISSORS | LFL | OLYMPUS MEDICAL SYSTEMS CORPORATION | T3955 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |