FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3770245
·
Received March 10, 2014
Report
- Report Number
- 1720753-2014-02142
- Event Type
- Malfunction
- Date Received
- March 10, 2014
- Date of Event
- February 18, 2014
- Report Date
- March 10, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE VOLTAGE ON POWER SUPPLY WAS EVALUATED AND ADJUSTED TO OPTIMUM OPERATING VOLTAGE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS A COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR IS LIKELY TO RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURRED. THERE IS NO REPORT OF PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.