FDA Adverse Event Malfunction Summary report: N

TRILOGY 200

MDR report key: 3770228 · Received March 12, 2014

Report

Report Number
2518422-2014-00373
Event Type
Malfunction
Date Received
March 12, 2014
Date of Event
February 13, 2014
Report Date
February 13, 2014
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K093416
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, "SERVICE REQUIRED" CODES WERE FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD AND INTERNAL BATTERY WERE REPLACED TO ADDRESS THE ISSUES.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149084 TRILOGY 200 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1040005

Patients

Seq Age Sex Outcome Treatment
1