FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3770223
·
Received March 10, 2014
Report
- Report Number
- 1720753-2014-02158
- Event Type
- Malfunction
- Date Received
- March 10, 2014
- Date of Event
- February 26, 2014
- Report Date
- March 10, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE HARD DISK DRIVE WAS EVALUATED AND REPLACED. THE SYSTEM SOFTWARE AND CALIBRATIONS WERE ALSO RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT TO A USABLE STATE AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143644 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |