FDA Adverse Event Injury Summary report: N

12/14 ARTICUL 40MM M SPEC+5

MDR report key: 3770167 · Received April 24, 2014

Report

Report Number
1818910-2014-17428
Event Type
Injury
Date Received
April 24, 2014
Date of Event
April 3, 2014
Report Date
August 8, 2014
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
LPH
PMA / PMN Number
PK060031
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PATIENT WAS REVISED TO ADDRESS PSEUDOTUMORS. INFECTION WAS ALSO REPORTED. UPDATE REC¿D 05/08/2014 - PATIENT'S MEDICAL RECORDS WERE RECEIVED. RECORDS INDICATE THE PSEUDOTUMORS THAT WERE PRESENT IN THE HIP WERE MOST LIKELY AN ADVERSE TISSUE REACTION TO THE METAL ON METAL HIP. DOI: (B)(6) 2010 - DOR: (B)(6) 2014 (RIGHT HIP). THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER RELATED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A. MEDICAL RECORDS WERE OBTAINED AND REVIEWED BY A MEDICAL PROFESSIONAL. FROM A MEDICAL PERSPECTIVE, BASED ON THE INFORMATION AVAILABLE, IT IS NOT POSSIBLE TO DETERMINE IF THE COMPLAINT IS PRODUCT RELATED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS PSEUDOTUMORS. INFECTION WAS ALSO REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249420 12/14 ARTICUL 40MM M SPEC+5 HIP FEMORAL HEAD LPH DEPUY ORTHOPAEDICS INC US 3072103

Patients

Seq Age Sex Outcome Treatment
1 58 YR Required Intervention