FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3770102
·
Received January 28, 2014
Report
- Report Number
- 1720753-2014-00943
- Event Type
- Malfunction
- Date Received
- January 28, 2014
- Date of Event
- January 22, 2014
- Report Date
- January 28, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE 5VDC POWER SUPPLY CONNECTIONS AND CONNECTORS WERE EVALUATED, CLEANED AND RESEATED. THE 5VDC POWER SUPPLY WAS ALSO CALIBRATED TO THE OPTIMAL OPERATING VOLTAGE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS A COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR IS LIKELY TO RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURRED. THERE IS NO REPORT OF A PATIENT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 61442 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |