FDA Adverse Event Malfunction Summary report: N

PLUMSET CLAVE PORT CLAVE SITE 103IN NDHP

MDR report key: 3770085 · Received January 28, 2014

Report

Report Number
9615050-2014-00723
Event Type
Malfunction
Date Received
January 28, 2014
Date of Event
January 1, 2013
Report Date
January 2, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K982159
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACT INDICATED THE DEVICES WERE DISCARDED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

GENERAL REPORT RECEIVED OF AN UNSPECIFIED NUMBER OF UNDOCUMENTED INCIDENTS OF DEVICE BREAKAGE; SUBSEQUENTLY, LEAKS WERE NOTED. ON UNSPECIFIED DATES, THE PRIMARY TUBING SETS WERE BEING USED TO DELIVER UNSPECIFIED MEDICATIONS, AT AN UNSPECIFIED RATE. AFTER UNSPECIFIED LENGTHS OF TIME, IT WAS REPORTED THAT THE MALE ADAPTER OF SECONDARY TUBING SETS WERE CONNECTED TO THE CLAVE SECONDARY PORT ON THE CASSETTE OF THE TUBING SETS FOR PIGGYBACK DELIVERIES OF UNSPECIFIED SOLUTIONS. AFTER UNSPECIFIED LENGTHS OF TIME AFTER THE PIGGYBACK DELIVERIES WERE STARTED, THE CUSTOMER CONTACT REPORTED THAT THE SEMI-RIGID ADAPTER OF THE CLAVE SECONDARY PORT ON THE CASSETTE OF THE TUBING SETS BROKE OFF AND UNSPECIFIED VOLUMES OF SOLUTION LEAKED. NO SPECIFIC DETAILS WERE PROVIDED. THE CUSTOMER CONTACT INDICATED THAT THE SEMI-RIGID ADAPTERS WERE FLIMSY AND BENT EASILY. THE TUBING SETS WERE REPLACED AND THE THERAPIES WERE RESUMED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO REPORTED DELAYS OF THERAPIES CRITICAL TO THESE PATIENTS. NO MEDICAL INTERVENTIONS WERE REPORTED. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
61082 PLUMSET CLAVE PORT CLAVE SITE 103IN NDHP 80FRN FRN HOSPIRA COSTA RICA LTD. NA UNK5H

Patients

Seq Age Sex Outcome Treatment
1 UNK SECONDARY TUBING SETS, LIST #UNK, LOT #UNK