FDA Adverse Event Malfunction Summary report: N

TX1 TISSUE REMOVAL SYSTEM - CONSOLE

MDR report key: 3769128 · Received December 30, 2013

Report

Report Number
2085033-2013-00167
Event Type
Malfunction
Date Received
December 30, 2013
Date of Event
November 5, 2012
Report Date
December 29, 2013
Manufacturer
AOI, INC
Product Code
LFL
PMA / PMN Number
K123640
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CONSOLE WAS RETURNED FOR EVALUATION. COULD NOT VERIFY THE MALFUNCTION. THE UNIT PRIMED SUCCESSFULLY. DURING THE EXECUTION OF PROTOCOL NUMBER P12-0048, IRRIGATION DID NOT SHUT OFF AFTER PROLONGED USAGE. IT WAS DETERMINED THAT THE VACUUM SENSOR DIAPHRAGM MOVEMENT IS BEING RESTRICTED, WHICH PREVENTS A COMPLETE RELEASE OF VACUUM CAUSING THE REFLUX VALVE TO REMAIN OPEN. TO CORRECT THE PROBLEM, THE SPACER VALVES WERE MANUFACTURED ON THE HIGHER END OF THEIR SPECIFICATION TO ENSURE CONTACT OF THE HANDPIECE CARTRIDGE WITH THE CONSOLE AND AN ELLIPTICAL FEATURE WAS ADDED TO THE CASSETTE TO ALLOW FOR THE SENSOR TO MOVE MORE FREELY. UNITS THAT WERE RETURNED THROUGH THE COMPLAINT SYSTEM WERE SERVICED AND THE SPACER VALVES WERE REPLACED PER CAPA (B)(4). THERE HAVE BEEN NO OTHER REPORTED ISSUES CONCERNING THIS CONSOLE. WHAT APPEARS TO BE FLUID BUILDUP, ONLY SEEMS TO OCCUR DURING AN ELBOW PROCEDURE DUE TO THE SKIN BEING TIGHT IN THIS REGION. THE COMPANY'S CHIEF MEDICAL OFFICER HAS DOCUMENTED THAT THERE IS NO HARM TO THE PATIENT WITH THE EXCESS FLUID. THE BODY WILL ABSORB IT. THERE HAVE BEEN NO REPORTED EVENTS OF FLUID RETENTION SINCE (B)(6) 2012. DOCTORS WITH MORE CASES TEND TO SQUEEZE OUT EXCESS FLUID AS PART OF THEIR ROUTINE. A PRODUCT RECALL WAS INITIATED MAY 2013 TO REPLACE SPACER VALVES ON CONSOLES IN DISTRIBUTION.

Description of Event or Problem · 1

SURGEON SAID A LOT OF FLUID REMAINED IN THE JOINT LEAVING A TENNIS BALL SIZE BUMP ON THE PATIENT'S ELBOW. THE SURGEON THEN PUSHED THE REMAINING AMOUNT OF FLUID OUT OF THE PATIENT'S ELBOW. HE ASKED THAT WE SEND THE CONSOLE BACK TO CHECK FOR ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681421 TX1 TISSUE REMOVAL SYSTEM - CONSOLE ULTRASONIC SURGICAL ASPIRATOR MGI LFL AOI, INC

Patients

Seq Age Sex Outcome Treatment
1