FDA Adverse Event Other Summary report: N

AFFUBUTT 4 BIRTHING BED

MDR report key: 3767703 · Received April 10, 2014

Report

Report Number
1824206-2014-01130
Event Type
Other
Date Received
April 10, 2014
Date of Event
March 11, 2014
Report Date
March 12, 2014
Manufacturer
HILL-ROM, INC.
Product Code
HDD
PMA / PMN Number
K915779
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH INVESTIGATED THE ALLEGED INCIDENT AND FOUND THAT WHEN THE FOOT SECTION WAS INSTALLED AND IF THE FOOT SECTION WAS SLID TO ONE SIDE IF IT WAS JUST TAPPED WITH A FINGER, THE FOOT SECTION FELL TO THE GROUND. THE SUSPECT FOOT SECTION MEASURED 20 1/4 INCHES AT THE SHOULDER OF THE BRACKETS AND 20 1/8 AT THE TIPS. THE RIGHT SIDE BRACKET APPEARED TO HAVE A SLIGHT BEND TO IT, IT WAS UNSURE IF THIS WAS A RESULT OF THE PT FALLING ON IT. THE HILL-ROM REP SPOKE WITH THE TECH TO CLARIFY THE MALFUNCTION AND THE TECH STATED THAT EVEN IF THE BRACKET WAS NOT BENT, THE FOOT SECTION IS NARROWER THAN THE YOKE AND DOES NOT ENGAGE THE RECEIVING BRACKETS. THE FOOT SECTION HAS BEEN SENT BACK TO HILL-ROM FOR FURTHER INVESTIGATION. THE HILL-ROM REP SPOKE WITH THE ACCOUNT'S VP OF QUALITY AND SHE ADVISED THE PT SUSTAINED MINOR SCRATCHES AND BRUISES. MEDICAL INTERVENTION WAS NOT REQUIRED. THE INCIDENT OCCURRED POST-DELIVERY OF THE BABY. EXTENDED HOSPITALIZATION WAS NOT REQUIRED. THE REPORTED INJURY IS NON-SERIOUS IN NATURE PER FDA DEFINITION. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNK IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE FOOT SECTION WAS REPLACED BY THE TECH TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THAT THE PT WAS TRYING TO GET IN THE BED BY THE FOOT SECTION AND THE FOOT SECTION DROPPED AND THE PT FELL TO THE FLOOR. THE BED WAS LOCATED IN THE LABOR AND DELIVERY AT THE FACILITY. (B)(6) ALLEGED THAT THE PT WAS INJURED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
216521 AFFUBUTT 4 BIRTHING BED BIRTHING BED HDD HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1 UNK Other