FDA Adverse Event Injury Summary report: N

MPACT HC PE FIXED FLAT LINER 36

MDR report key: 3764732 · Received April 9, 2014

Report

Report Number
3005180920-2014-00040
Event Type
Injury
Date Received
April 9, 2014
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
PMA / PMN Number
K103721
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

DOCUMENT REVIEW: MPACT HC UHMWPE ACETABULAR LINER SIZE 36E: REF. 01.32.3644HCT/ LOT 132131 ((B)(4) LINERS PRODUCED): ALL PARAMETERS WERE FOUND TO BE IN ACCORDANCE WITH THE SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING, INCLUDING WASHING AND STERILIZATION PROCEDURES. (B)(4) LINERS BELONGING TO THIS LOT HAVE BEEN ALREADY IMPLANTED WITHOUT ANY OTHER SIMILAR INCIDENT. MPACT ACETABULAR SHELL TWO HOLES SIZE 50: REF. 01.32.050DH/ LOT 132790 ((B)(4) SHELLS PRODUCED): ALL PARAMETERS WERE FOUND TO BE IN ACCORDANCE WITH THE SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING, INCLUDING WASHING AND STERILIZATION PROCEDURES. (B)(4) CUPS BELONGING TO THIS LOT HAVE BEEN ALREADY IMPLANTED WITHOUT ANY OTHER SIMILAR INCIDENT. FROM THE INFORMATION COLLECTED, THE EVENT IS HIGHLY LIKELY NOT DEVICE RELATED.

Description of Event or Problem · 1

REFERENCE IMPORTER # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
213312 MPACT HC PE FIXED FLAT LINER 36 ACETABULAR PE LINER LZO MEDACTA INTERNATIONAL SA

Patients

Seq Age Sex Outcome Treatment
1