FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3761311 · Received April 21, 2014

Report

Report Number
2531779-2014-10686
Event Type
Malfunction
Date Received
April 21, 2014
Report Date
April 6, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 05/20/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/12/2014 WITH THE FOLLOWING FINDINGS:A REVIEW OF THE PUMP BLACK BOX SHOWED MULTIPLE UNEXPLAINED POWER EVENTS. THE PUMP BATTERY COMPARTMENT AND CAP WERE OBSERVED TO BE INTACT. THE BATTERY CAP CONTACTS WERE MEASURED AND CONFIRMED TO BE WITHIN SPECIFICATIONS. THE BATTERY CAP WAS FOUND TO BE ABLE TO FULLY TIGHTEN TO THE PUMP. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT POWER ISSUES OR ALARMS. THE PUMP WAS OPENED AND THERE WAS NO EVIDENCE OF INTERNAL MOISTURE OR LOOSE COMPONENTS WITHIN THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER EVENT WHICH WAS RESOLVED BY REPLACING THE PUMP BATTERY. A REVIEW OF THE PUMP HISTORY FOUND NO INDICATION OF A BATTERY OR OTHER ALARM RELATED TO THE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
240302 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 48 YR