FDA Adverse Event Malfunction Summary report: N

R SERIES DEFIBRILLATOR

MDR report key: 3761269 · Received January 2, 2014

Report

Report Number
1220908-2013-03553
Event Type
Malfunction
Date Received
January 2, 2014
Report Date
December 17, 2013
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K060559
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY WAS DISTORTED. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1112 R SERIES DEFIBRILLATOR DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION R SERIES NA

Patients

Seq Age Sex Outcome Treatment
1 NA