STOCKERT EP-SHUTTLE RADIO FREQUENCY GENERATOR
Report
- Report Number
- 9612355-2014-00021
- Event Type
- Malfunction
- Date Received
- April 21, 2014
- Date of Event
- March 26, 2014
- Report Date
- March 31, 2014
- Manufacturer
- STOCKERT GMBH
- Product Code
- DRF
- PMA / PMN Number
- P990071
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). IT WAS REPORTED THAT DURING AN ATRIAL FIBRILLATION (AFIB) PROCEDURE, THE DISPLAYED TEMPERATURE OF THE SMART TOUCH CATHETER WAS 53 DEGREE CELSIUS. THE ABLATION COULD BE CONDUCTED AT 53 DEGREE CELSIUS ALTHOUGH THE CUT-OFF WAS SET TO 44 DEGREE CELSIUS. AS THE PHYSICIAN NOTICED THE ISSUE, HE STOPPED THE ABLATION. AFTER THAT, THE TEMPERATURE CUT-OFF FUNCTION WAS ACTIVATED PROPERLY SPONTANEOUSLY AS THE TEMPERATURE DID NOT SOAR UNTIL 44 DEGREE CELSIUS. THE PROCEDURE WAS COMPLETED WITH USING THE CATHETER. THE PROCEDURE WAS COMPLETED WITHOUT PATIENT CONSEQUENCE. THE DEVICE WAS EVALUATED AND THE ISSUE WAS NOT REPRODUCIBLE. THE SYSTEM WAS REPLACED FOR PREVENTION PURPOSES. REPLACEMENT OF THE MODULE 331 WAS FOR PREVENTIVE PURPOSES. THE DEVICE HISTORY RECORD WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES.
INVESTIGATION STILL IN PROGRESS. A SUPPLEMENTAL REPORT OR DEVICE EVALUATION WILL BE SUBMITTED. CONCOMITANT PRODUCT: SMART TOUCH BIDIRECTIONAL, MFG #: D-1327-05-S, LOT #: 16040760M. (B)(4).
IT WAS REPORTED THAT DURING AN ATRIAL FIBRILLATION (AFIB) PROCEDURE, THE DISPLAYED TEMPERATURE OF THE SMART TOUCH CATHETER WAS 53 DEGREE CELSIUS. THE ABLATION COULD BE CONDUCTED AT 53 DEGREE CELSIUS ALTHOUGH THE CUT-OFF WAS SET TO 44 DEGREE CELSIUS. AS THE PHYSICIAN NOTICED THE ISSUE, HE STOPPED THE ABLATION. AFTER THAT, THE TEMPERATURE CUT-OFF FUNCTION WAS ACTIVATED PROPERLY SPONTANEOUSLY AS THE TEMPERATURE DID NOT SOAR UNTIL 44 DEGREE CELSIUS. THE PROCEDURE WAS COMPLETED WITH USING THE CATHETER. THE PROCEDURE WAS COMPLETED WITHOUT PATIENT CONSEQUENCE. PER THE EVENT DESCRIPTION STATING, "THE ABLATION COULD BE CONDUCTED AT 53 DEGREES CELSIUS ALTHOUGH THE CUT-OFF WAS SET TO 44 DEGREE CELSIUS" IS INDICATIVE OF A REPORTABLE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 240567 | STOCKERT EP-SHUTTLE RADIO FREQUENCY GENERATOR | RADIOFREQUENCY GENERATOR AND ACCESSORIES/ACCESSORY CABLE | DRF | STOCKERT GMBH | 39D-76X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |