COBAS INTEGRA 400 PLUS
Report
- Report Number
- 1823260-2014-02873
- Event Type
- Malfunction
- Date Received
- April 21, 2014
- Date of Event
- March 29, 2014
- Report Date
- May 5, 2014
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K951595
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE SPECIFIC ROOT CAUSE OF THE POWER BOX FAILURE COULD NOT BE DETERMINED. PICTURES OF THE AFFECTED POWER BOX WERE PROVIDED AND IT CAN BE SEEN THAT PART OF THE PLANAR WINDING WAS MELTED. WITH A HIGH PROBABILITY, THIS CAUSED THE SMOKE COMING FROM THE INSTRUMENT. THERE WAS NO IMMINENT DANGER OF FIRE RELATED TO THIS POWER BOX FAILURE. ALL PARTS AND PLASTICS OF THE POWER BOX ARE UL RATED ACCORDINGLY.
IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.
THE CUSTOMER REPORTED THAT THERE WAS SMOKE COMING FROM THE ANALYZER. THE ANALYZER WAS SHUT DOWN AND THE LABORATORY WAS EVACUATED. THE LABORATORY WAS FULL OF SMOKE AND A "BOTHERSOME FUME." THE FIRE DEPARTMENT ARRIVED AND THE LABORATORY PERSONNEL DID NOT GO BACK INSIDE THE LABORATORY UNTIL THE FIRE DEPARTMENT GAVE THE CLEAR. A FEMALE PHLEBOTOMIST IN THE LABORATORY LATER COMPLAINED OF A HEADACHE AS A RESULT OF SMOKE INHALATION. THE OFFICE NURSE GAVE THIS PERSON AN OVER THE COUNTER PAIN RELIEVER AS TREATMENT. THE PAIN RELIEVER RESOLVED THE HEADACHE AND THERE WAS NO FURTHER INCIDENT. IT WAS ASKED, BUT NO FURTHER INFORMATION COULD BE PROVIDED ABOUT THE PHLEBOTOMIST. NO OTHER HARM WAS EXPERIENCED BY THE LABORATORY PERSONNEL. THE FIELD SERVICE REPRESENTATIVE DETERMINED THAT THE POWER SUPPLY FAN FAILED, CAUSING THE POWER SUPPLY TO OVER HEAT. HE REPLACED THE POWER SUPPLY, WIRE HARNESS, AND POWER CONTROLLER PCB. HE BOOTED UP THE INSTRUMENT AND PERFORMED VOLTAGE MEASUREMENTS. HE RAN ALL AUTO MAINTENANCE FUNCTIONS SUCCESSFULLY. HE COMPLETED ISE CALIBRATIONS. THE OPERATOR RAN CALIBRATION AND QUALITY CONTROLS WITH ALL ASSAYS RECOVERING WITHIN SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 241173 | COBAS INTEGRA 400 PLUS | CLINICAL CHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |