FDA Adverse Event
Malfunction
Summary report: N
CD HORIZON SPINAL SYSTEM
MDR report key: 3761162
·
Received April 21, 2014
Report
- Report Number
- 1030489-2014-02248
- Event Type
- Malfunction
- Date Received
- April 21, 2014
- Date of Event
- March 21, 2014
- Report Date
- March 23, 2014
- Manufacturer
- MDT SOFAMOR DANEK PUERTO RICO MFG
- Product Code
- KWP
- PMA / PMN Number
- K981676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ADDITIONAL INFO: MACROSCOPIC AND OPTICAL EXAMINATION REVEALED THREAD CREST AND FLANK DAMAGE; THIS DAMAGE APPEARS TO HAVE INITIATED AT THE START OF THE THREAD, AND IS CONSISTENT AROUND THE THREAD. THE ABOVE OBSERVATIONS ARE CONSISTENT WITH MISALIGNMENT OF THE MAS AND SET SCREW THREADS DURING CONSTRUCT ASSEMBLY.
Additional Manufacturer Narrative · 1
(B)(4): NEITHER THE DEVICE NOR APPLICABLE IMAGING FILMS WERE RETURNED TO THE MANUFACTURER FOR EVALUATION, THEREFORE, THE CAUSE OF THE EVENT CANNOT BE DETERMINED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT UNDERWENT A POSTERIOR SPINAL SURGICAL PROCEDURE TO TREAT LUMBAR SPINAL STENOSIS. IT WAS REPORTED THAT THE SET SCREW SLIPPED DURING TIGHTENING IN THE BONE SCREW. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 240519 | CD HORIZON SPINAL SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | MDT SOFAMOR DANEK PUERTO RICO MFG | NA | H13E2974 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00061 YR | BONE SCREW |