FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3761142
·
Received December 31, 2013
Report
- Report Number
- 3008642652-2013-03743
- Event Type
- Malfunction
- Date Received
- December 31, 2013
- Date of Event
- December 3, 2013
- Report Date
- December 31, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON INVESTIGATION, THE DISTRIBUTION NODE (DN) TO REAR THERAPY ELECTRODE (TE) CABLE WAS PULLED FROM THE STRAIN RELIEF. THE DISCONNECTED CABLE CAUSED THE BELT TO FAIL THE LEAD HI-POT TEST. THE ROOT CAUSE OF THE DAMAGED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING, ELECTRODE BELT SN (B)(4), FAILED THE HI-POT TEST. THE LAST PATIENT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 682951 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |