ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2014-02727
- Event Type
- Malfunction
- Date Received
- April 21, 2014
- Date of Event
- March 31, 2014
- Report Date
- April 4, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION IS UNAVAILABLE. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. AT THE TIME OF THIS SUBMISSION, THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING 6 CLIPS AS INTENDED. IN ORDER TO CONFIRM THE CLIPS WERE WITHIN MANUFACTURING SPECIFICATIONS, THE CLIPS WERE EVALUATED USING A TOOL THAT IS DESIGNED TO DETERMINE PROPER FORMATION. DURING ANALYSIS THE JAWS OPENED AND CLOSED AS INTENDED. IN ADDITION THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THREE CLIPS WERE APPLIED TO THE DUCT AND WOULD NOT HOLD ONTO THE TISSUE. THE CLIPS SLIPPED OFF THE TISSUE. THE CASE WAS COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. NO CHOLANGIOGRAM WAS USED WITH THIS PROCEDURE. ALL AVAILABLE INFORMATION WAS REPORTED. NO PATIENT CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 241134 | ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | K4DR7W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |