FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 3761120 · Received December 31, 2013

Report

Report Number
3004464228-2013-01620
Event Type
Malfunction
Date Received
December 31, 2013
Date of Event
December 4, 2013
Report Date
December 4, 2013
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K122953
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE CUSTOMER REPORTED THAT THE CANNULA HAD DISLODGED FROM THE INFUSION SITE, A CONDITION WHICH COULD INTERRUPT INSULIN DELIVERY AND CONTRIBUTE TO HYPERGLYCEMIA QUALIFICATION RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD'S USER GUIDE WARNS "CHECK OFTEN TO MAKE SURE THE POD AND SOFT CANNULA ARE SECURELY ATTACHED AND IN PLACE. A LOOSE OR DISLODGED CANNULA MAY INTERRUPT DELIVERY VERIFY THERE IS NO WETNESS OR SCENT OF S=INSULIN, WHICH MAY INDICATE THE CANNULA HAS DISLODGED," BECAUSE INSULIN PODS USE ONLY RAPID-ACTING INSULIN, USERS ARE AT INCREASED RISK FOR DEVELOPING HYPERGLYCEMIA (HIGH BLOOD GLUCOSE) IF INSULIN DELIVERY IS INTERRUPTED," AND "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER."

Description of Event or Problem · 1

THE CUSTOMER REPORTED ON THE SECOND DAY AFTER THE POD WAS APPLIED ON HER ARM HER BLOOD GLUCOSE REACHED OVER 250 MG/DL. SHE REALIZED THE CANNULA WAS NOT IN THE SKIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
682513 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 14810-5E L40582

Patients

Seq Age Sex Outcome Treatment
1