ANGIOSCULPT PTCA SCORING BALLOON CATHETER (RX)
Report
- Report Number
- 3005462046-2014-00010
- Event Type
- Injury
- Date Received
- April 7, 2014
- Date of Event
- March 10, 2014
- Report Date
- March 10, 2014
- Manufacturer
- ANGIOSCORE, INC.
- Product Code
- NWX
- PMA / PMN Number
- P050018
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AN AIR EMBOLISM OCCURRED DURING USE OF THE ANGIOSCULPT DEVICE. THE PT WAS GIVEN MEDICATION, THUS RESULTING IN ADDITIONAL INTERVENTION. THE ANGIOSCULPT DEVICE WAS RETURNED FOR EVALUATION. VISUAL EXAMINATION CONFIRMED THE BALLOON HAD BEEN INFLATED AND THE RX PORT IS LACERATED. DURING FUNCTIONAL TESTING, THE BALLOON WAS INFLATED TO 2 ATM AND A PINHOLE LEAK WAS OBSERVED DISTAL TO THE PROXIMAL MARKER BAND. PER THE ANGIOSCULPT IFU, EMBOLISM IS LISTED AS A POSSIBLE ADVERSE EFFECT OF THE PROCEDURE.
THE PT HAD IN-STENT RESTENOSIS. THE PHYSICIAN ATTEMPTED TO BALLOON THE LESION WITH AN ANGIOSCULPT BALLOON. THE BALLOON WAS INFLATED TWICE TO 8 ATM FOR 30 SECONDS. THE BALLOON WAS THEN REMOVED OUT OF THE BODY AND AN ANGIOGRAM WAS PERFORMED AND SHOWED THE LESION WAS STILL PRESENT. THE BALLOON WAS REINTRODUCED TO THE RCA AND WAS INFLATED TO 12 ATM. THE INDEFLATOR WAS UNABLE TO HOLD PRESSURE SO THE DEVICE WAS DEFLATED AND REMOVED FROM THE PT. AN ANGIOGRAM WITHOUT CONTRAST INJECTED AFTER SHOWED CONTRAST IN THE VESSEL AND AN AIR EMBOLISM. THE PT WENT BRADYCARDIC AND HER BLOOD PRESSURE DROPPED. THE EKG SHOWED ST ELEVATION. THE PT WAS GIVEN MEDICATIONS FORM THE NURSE AND WAS STABILIZED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 205521 | ANGIOSCULPT PTCA SCORING BALLOON CATHETER (RX) | CATHETER, ANGIOPLASTY, CORONARY | NWX | ANGIOSCORE, INC. | 2200-3015 | F13090067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Other| R | 0.014" GUIDE WIRE/6F INTRODUCER SHEATH/6F GUIDE| CATHETER |