FDA Adverse Event
Malfunction
Summary report: N
ADJ PIN COLLET 2.0-3.2MM
MDR report key: 3760765
·
Received April 21, 2014
Report
- Report Number
- 0001811755-2014-01397
- Event Type
- Malfunction
- Date Received
- April 21, 2014
- Date of Event
- March 24, 2014
- Report Date
- March 24, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY, DEBRIS WAS DISCOVERED NEAR THE COLLET WEDGE AND SMALL AMOUNTS OF CORROSION WERE FOUND INTERNALLY. THE PRESENCE OF CORROSION AND DEBRIS CAN CAUSE OR CONTRIBUTE TO METAL SHAVINGS COMING FROM THE DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT UNKNOWN PIECES OF BLACK MATERIAL WERE OBSERVED TO COME FROM THE ADJ PIN COLLET 2.0-3.2MM DURING CLEANING. THERE WAS NO PATIENT INVOLVEMENT, AND NO USER INJURIES OR ADVERSE CONSEQUENCES.