FDA Adverse Event Malfunction Summary report: N

ADJ PIN COLLET 2.0-3.2MM

MDR report key: 3760765 · Received April 21, 2014

Report

Report Number
0001811755-2014-01397
Event Type
Malfunction
Date Received
April 21, 2014
Date of Event
March 24, 2014
Report Date
March 24, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY, DEBRIS WAS DISCOVERED NEAR THE COLLET WEDGE AND SMALL AMOUNTS OF CORROSION WERE FOUND INTERNALLY. THE PRESENCE OF CORROSION AND DEBRIS CAN CAUSE OR CONTRIBUTE TO METAL SHAVINGS COMING FROM THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT UNKNOWN PIECES OF BLACK MATERIAL WERE OBSERVED TO COME FROM THE ADJ PIN COLLET 2.0-3.2MM DURING CLEANING. THERE WAS NO PATIENT INVOLVEMENT, AND NO USER INJURIES OR ADVERSE CONSEQUENCES.