FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL WITH HANDSWITCH

MDR report key: 3760721 · Received April 21, 2014

Report

Report Number
0001811755-2014-01386
Event Type
Malfunction
Date Received
April 21, 2014
Date of Event
March 24, 2014
Report Date
March 24, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING FAILURE ANALYSIS, THE REPORTED EVENT OF THE DEVICE RUNNING IN SAFE MODE WAS CONFIRMED. THE DEVICE HAD A LOOSE LEVER ASSEMBLY DUE TO LOCTITE FAILURE. A LOOSE LEVER ASSEMBLY CAN PREVENT THE SAFETY SWITCH FROM BEING ENGAGED, CAUSING THE DEVICE TO HAVE RUN ON.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS. A FOLLOW-UP WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETE. DEVICE EVALUATION IN PROGRESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH RAN IN SAFE MODE DURING SERVICE EVALUATION AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MAESTRO DRILL WITH HANDSWITCH RAN IN SAFE MODE DURING SERVICE EVALUATION AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, NO ADVERSE CONSEQUENCES, NO MEDICAL INTERVENTION, AND NO SURGICAL DELAY WERE REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
239338 MAESTRO DRILL WITH HANDSWITCH MOTOR, DRILL, PNEUMATIC HBB STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1