FDA Adverse Event Death Summary report: N

ACL TOP 500

MDR report key: 3760662 · Received April 11, 2014

Report

Report Number
1217183-2014-00003
Event Type
Death
Date Received
April 11, 2014
Date of Event
March 9, 2014
Report Date
March 13, 2014
Manufacturer
INSTRUMENTATION LABORATORY CO.
Product Code
GKP
PMA / PMN Number
K07337
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN INTERNAL EVALUATION HAS BEEN PERFORMED WITH THE REAGENT LOT OF HEMOSIL Q.F.A. THROMBIN IN USE AT THE CUSTOMER SITE. THE EVALUATION SHOWED THAT THE TEST SUCCESSFULLY CALIBRATED AND RECOVERED MEANS FOR ALL CONTROL SETS WITHIN THEIR RANGES. TO COMPLETE THE INVESTIGATION, INSTRUMENTATION LABORATORY CO. IS AWAITING THE DATA BACK-UP FROM THE ACL TOP 500 CTS IN USE AT THE CUSTOMER SITE. SEVERAL REQUESTS HAVE BEEN MADE FOR THE BACK-UP AND WE WILL SUBMIT A FOLLOW-UP REPORT WHEN THIS ADDITIONAL INFO IS RECEIVED. THERE IS NO INDICATION TO DATE THAT THE HEMOSIL Q.F.A. THROMBIN RESULT WAS A CONTRIBUTING FACTOR IN THE PT DEATH.

Description of Event or Problem · 1

ON (B)(6) 2014, THE CUSTOMER CALLED TO QUESTION A PT'S CLAUSS FIBRINOGEN RESULT USING HEMOSIL Q.F.A. THROMBIN ON THEIR ACL TOP 500 CTS INSTRUMENT. THE SAMPLE WAS DRAWN ON (B)(6) 2014, AND GAVE A FIBRINOGEN RESULT OF 183 MG/DL. DAILY MAINTENANCE WAS PERFORMED ON THE INSTRUMENT AND QC WAS IN RANGE BEFORE AND AFTER THE SAMPLE WAS DRAWN. THE SAMPLE WAS DRAW AT 12:10 PM AND FROZEN AT 2:00 PM. THE FROZEN SAMPLE WAS DEFROSTED FOUR DAYS POST DRAW AND THAWED AT 37 DEGREES CELSIUS FOR FOUR MINUTES. THE THAWED, DILUTED, SAMPLE WAS RUN IN THE Q/F/A/ TEST AT 11/35 ON 03/13 AND FAILED. TI WAS THEN TESTED AGAIN AS A NEAT SAMPLE AT 11:36 AND A RESULT OF 332 MG/DL WAS OBTAINED. THE SAMPLE WAS THEN TESTED IN THE Q.F.A. LOW TEST AND A RESULT OF 67 MG/DL WAS OBTAINED. THE PT DIED ON (B)(6) 2014. THERE IS NO INDICATION FROM THE CUSTOMER THAT THE FIBRINOGEN RESULT OF 183 MG/DL WAS RELATED TO THE PT DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
220662 ACL TOP 500 COAGULATION ANALYZER GKP INSTRUMENTATION LABORATORY CO. 2800-40 NA

Patients

Seq Age Sex Outcome Treatment
1 Death