ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-10656
- Event Type
- Malfunction
- Date Received
- April 21, 2014
- Report Date
- April 8, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 05/16/2014. DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/08/2014 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE PUMP POWERED ON WITH NO DISPLAY. THE AUDIBLE AND VIBRATORY FEATURES FUNCTIONED APPROPRIATELY. THE PUMP WAS OPENED AND THE DISPLAY WAS FOUND TO BE CRACKED AND DAMAGED. WHEN THE DISPLAY WAS REPLACED BY A TEST DISPLAY, THE DISPLAY FUNCTIONED PROPERLY. UNRELATED TO THE DISPLAY ISSUE, EVALUATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.