ANIMAS VIBE
Report
- Report Number
- 2531779-2014-10584
- Event Type
- Malfunction
- Date Received
- April 21, 2014
- Report Date
- March 27, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4). DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE DISPLAY SCREEN WAS BLANK WHEN THE PUMP STARTED UP BUT THE PUMP DID EMIT AUDIO TONES. THE DISPLAY SCREEN WAS FOUND TO BE CRACKED. THE DAMAGED DISPLAY WAS REPLACED WITH A TEST SCREEN AND THE TEST SCREEN FUNCTIONED NORMALLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, A REPORTER CONTACTED ANIMAS ALLEGING THAT THEIR DISPLAY SCREEN HAD BECOME DIM AND DISCOLORED. THIS COMPLAINT IS BEING REPORTED BECAUSE IT WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 240865 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 8 YR |