FDA Adverse Event Malfunction Summary report: N

PERFORATOR DRIVER W/HUDSON END

MDR report key: 3760264 · Received April 19, 2014

Report

Report Number
1045834-2014-11815
Event Type
Malfunction
Date Received
April 19, 2014
Report Date
March 27, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
PK042783
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INITIAL REPORTER: CONTACT NUMBER (B)(6). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DRIVE SHAFT WAS DAMAGED. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO MISUSE BY NOT EFFECTIVELY USING THE PROTECTIVE CAP THAT WAS PROVIDED TO GUARD THE SHAFT FROM DAMAGE WHILE NOT IN USE, WHICH IS MISUSE, ABUSE AND/OR USER ERROR. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED FROM (B)(6) THAT THE COMPACT SPEED REDUCER DEVICE WAS HEATING UP WHEN IN USE. IT WAS UNKNOWN TO THE REPORTER IF THE DEVICE WAS USED IN SURGERY. IT WAS UNKNOWN TO THE REPORTER IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS UNKNOWN IF INJURIES OR MEDICAL INTERVENTION WERE REPORTED. THE DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
239095 PERFORATOR DRIVER W/HUDSON END DRILLS, BURRS, TREPHINES & ACCESSORIES - COMPACT SPEED REDUCER HBE DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1