FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3760259 · Received April 19, 2014

Report

Report Number
1416980-2014-12742
Event Type
Malfunction
Date Received
April 19, 2014
Date of Event
January 17, 2014
Report Date
March 26, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED FOR THE REPORTED EVENT. A REVIEW OF THE EVENT HISTORY LOG CONFIRMED THE OCCURRENCE OF AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE THAT ONE OR MORE CYCLES ADVANCED TO THE NEXT FILL WHEN SLOW/NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 22:15:49. DURING NIGHT DRAIN CYCLE 25, THE PATIENT'S ULTRAFILTRATION READING WAS 1729ML, INDICATING THE HOME PATIENT (HP) DRAINED 979ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500MLNO ADDITIONAL INFORMATION