FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3760256 · Received April 19, 2014

Report

Report Number
1416980-2014-12739
Event Type
Malfunction
Date Received
April 19, 2014
Date of Event
March 4, 2014
Report Date
March 25, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS EVALUATED AND THE EVENT WAS CONFIRMED THROUGH REVIEW OF THE EVENT HISTORY LOGS. THE CAUSE OF THE REPORTED ISSUE WAS DETERMINED TO BE INSUFFICIENT DRAIN DUE TO ONE OR MORE CYCLES ADVANCING TO THE NEXT FILL WHEN SLOW / NO FLOW OCCURRED ABOVE THE MINIMUM DRAIN VOLUME THRESHOLD. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED DURING THERAPY INITIATED ON (B)(6) 2014 AT 12:17:05. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1312ML, INDICATING THE HOME PATIENT (HP) DRAINED 1312ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO FURTHER INFORMATION WAS AVAILABLE.