FDA Adverse Event Malfunction Summary report: N

ACCESS® 2 IMMUNOASSAY SYSTEM

MDR report key: 3760225 · Received April 18, 2014

Report

Report Number
2122870-2014-00275
Event Type
Malfunction
Date Received
April 18, 2014
Date of Event
March 21, 2014
Report Date
March 26, 2014
Manufacturer
BECKMAN COULTER
Product Code
MMI
PMA / PMN Number
K922823/A007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) ADJUSTED THE MIXER SPEED FROM 2,450 RPM TO 2,500 RPM (ROTATIONS PER MINUTE). THE FSE NOTED THE ULTRASONIC VOLTAGE WAS AT 228 VOLTS AND ADJUSTED IT TO 197 VOLTS (INCORRECT VOLTAGE MAY CAUSE INSUFFICIENT MIXING OF THE MAGNETIC PARTICLES). THE FSE PERFORMED A PASSING SYSTEM CHECK, HIGH SENSITIVITY SYSTEM CHECK, AND 15-REPLICATE PRECISION TEST. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. IN CONCLUSION, SYSTEM HARDWARE, ASSOCIATED WITH THE TRANSDUCER VOLTAGE, IS THE LIKELY CAUSE OF THE EVENT. BECKMAN COULTER CONTINUES TO TRACK AND TREND ANY INCIDENT RELATED TO THIS ISSUE. ALL RELATED MEDWATCH REPORTS ASSOCIATED WITH THIS INCIDENT: 2122870-2014-00274; 2122870-2014-00275; 2122870-2014-00276; 2122870-2014-00277.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FAILED CALIBRATIONS AND IND (INDETERMINATE) FLAGGED ACCESS® ACCUTNI+3 RESULTS, FOR FIVE PATIENTS, ON SEPARATE EVENT DATES, INVOLVING THE ACCESS® 2 IMMUNOASSAY SYSTEM. THE IND RESULTS (NON-NUMERICAL) WERE NOT RELEASED FROM THE LABORATORY. THE CUSTOMER REANALYZED THE PATIENTS' SAMPLES ON THE ALTERNATE ACCESS® 2 INSTRUMENT AND OBTAINED LOWER RESULTS. THERE WAS NO PATIENT INJURY OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236881 ACCESS® 2 IMMUNOASSAY SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE MMI BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1