FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3760185 · Received April 18, 2014

Report

Report Number
3004209178-2014-83805
Event Type
Injury
Date Received
April 18, 2014
Date of Event
April 2, 2014
Report Date
April 3, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED ALL FUNCTIONAL TEST, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. NO MOTOR ANOMALY NOTED OR RATTLES NOISE DURING REWIND. UNIT HAD SCRATCHED DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS TAKEN TO THE EMERGENCY ROOM AND HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE. THE BLOOD GLUCOSE READING WAS 643 MG/DL. CALLER STATED THAT THE CUSTOMER'S INSULIN PUMP HAS A RATTLING LOUD SOUND AT REWIND. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE HIGH-PRESSURE TEST TWICE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237046 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 59 YR Hospitalization