FDA Adverse Event Malfunction Summary report: N

FOLFUSOR

MDR report key: 3760148 · Received April 18, 2014

Report

Report Number
1416980-2014-12725
Event Type
Malfunction
Date Received
April 18, 2014
Date of Event
March 15, 2014
Report Date
March 25, 2014
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RECEIVED FOR EVALUATION WITH 18ML OF RESIDUAL FLUID IN ITS BLADDER. ACCORDING TO THE LABEL AFFIXED TO THE DEVICE, THE DEVICE HAD BEEN FILLED WITH FLUOROURACIL AND NORMAL SALINE. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. A FUNCTIONAL FLOW RATE TEST WAS PERFORMED, AND THE FLOW RATE OF THE DEVICE WAS FOUND TO BE WITHIN SPECIFICATION. THE REPORTED CONDITION WAS NOT CONFIRMED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). (B)(6). THE LOT WAS MANUFACTURED FROM 05/01/2013 TO 05/02/2013. THE DEVICE WAS RETURNED AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SMALL VOLUME FOLFUSOR EXPERIENCED AN UNDERINFUSION. THE EVENT WAS DESCRIBED AS "SLOW FLOW". THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237536 FOLFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 13E002

Patients

Seq Age Sex Outcome Treatment
1