FDA Adverse Event Malfunction Summary report: N

SAGITTAL SAW ATTACHMENT

MDR report key: 3759592 · Received April 18, 2014

Report

Report Number
2520274-2014-10648
Event Type
Malfunction
Date Received
April 18, 2014
Report Date
March 20, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HWE
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. IT WAS OBSERVED THAT THE BALL BEARINGS INSIDE THE DEVICE WERE DAMAGED. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMMERSION DURING CLEANING; USER ERROR/MISUSE OR ABUSE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

(B)(6). THE MANUFACTURER LOCATION WAS DOCUMENTED AS UNKNOWN IN THE INITIAL REPORT. THE LOCATION HAS BEEN UPDATED TO WALDENBURG. CONTACT OFFICE NAME/ADDRESS HAS BEEN UPDATED ACCORDINGLY TO REFLECT THE CORRECTED MANUFACTURING FACILITY. DEVICE MANUFACTURE DATE: THE DEVICE MANUFACTURE DATE WAS DOCUMENTED AS UNKNOWN IN THE INITIAL REPORT. THE DEVICE MANUFACTURE DATE HAS BEEN UPDATED AS JUN 18, 2010. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT FROM (B)(6) DURING A FEMORAL HEAD OSTECTOMY (FHO) VETERINARY SURGERY, IT WAS DISCOVERED THAT THE SAGITTAL SAW ATTACHMENT DEVICE WAS FROZEN AND WAS NOT FUNCTIONING. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237114 SAGITTAL SAW ATTACHMENT INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE DEPUY SYNTHES POWER TOOLS 11082

Patients

Seq Age Sex Outcome Treatment
1