FDA Adverse Event
Malfunction
Summary report: N
NEPTUNE 2 ROVER ULTRA (120V)
MDR report key: 3757921
·
Received April 17, 2014
Report
- Report Number
- 0001811755-2014-01380
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- March 20, 2014
- Report Date
- March 21, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- JCX
- PMA / PMN Number
- K132671
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FIELD SERVICE TECHNICIAN WAS ABLE TO CONFIRM THE REPORTED EVENT. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO THE AC POWER BOARD. A FAILURE OF AN ELECTRICAL COMPONENT CAN RESULT IN SPARKING AND A BURNING ODOR. THE TECHNICIAN REPLACED THE AC POWER BOARD AND THE ROVER FUNCTIONED AS INTENDED.
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS IS ONGOING; ADDITIONAL INFORMATION MAY BE SUBMITTED ONCE THE RESULTS ANALYSIS IS COMPLETE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE NEPTUNE 2 ROVER ULTRA WAS SPARKING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE NEPTUNE 2 ROVER ULTRA WAS SPARKING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.