FDA Adverse Event Malfunction Summary report: N

NEPTUNE 2 ROVER ULTRA (120V)

MDR report key: 3757921 · Received April 17, 2014

Report

Report Number
0001811755-2014-01380
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 20, 2014
Report Date
March 21, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
JCX
PMA / PMN Number
K132671
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE TECHNICIAN WAS ABLE TO CONFIRM THE REPORTED EVENT. THE ROOT CAUSE WAS DETERMINED TO BE DUE TO THE AC POWER BOARD. A FAILURE OF AN ELECTRICAL COMPONENT CAN RESULT IN SPARKING AND A BURNING ODOR. THE TECHNICIAN REPLACED THE AC POWER BOARD AND THE ROVER FUNCTIONED AS INTENDED.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS ONGOING; ADDITIONAL INFORMATION MAY BE SUBMITTED ONCE THE RESULTS ANALYSIS IS COMPLETE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NEPTUNE 2 ROVER ULTRA WAS SPARKING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE NEPTUNE 2 ROVER ULTRA WAS SPARKING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.