FDA Adverse Event Injury Summary report: N

WESTERN MEDICA MEDICAL OXYGEN PRESET REGULATOR

MDR report key: 3753560 · Received January 10, 2014

Report

Report Number
1526809-2014-00001
Event Type
Injury
Date Received
January 10, 2014
Date of Event
July 30, 2012
Report Date
December 13, 2013
Manufacturer
WESTERN/SCOTT FETZER CO.
Product Code
CAN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE M1-540-P IS A SINGLE STAGE, PRESET OXYGEN REGULATOR WITH MAXIMUM INLET PRESSURE OF 3000 PSI. THE OUTLET PRESSURE IS FACTORY PRESET TO DELIVER OUTLET PRESSURE WHEN THE SUPPLY (I.E., TANK) IS TURNED ON (I.E. DELIVERS 50-56 PSI AT A SUPPLY PRESSURE OF 2000-2200 PSI). THE MODEL HAS ONE, 2" DIAMETER GAUGE (I.E., BOURDON TUBE GAS PRESSURE GAUGE), WHICH DISPLAYS THE GAS PRESSURE SUPPLIED TO THE REGULATOR. THE CUSTOMER REPORTED THAT THE "TANK WAS EMPTY DESPITE REGULATOR GAUGE REFLECTING APPROXIMATELY 800-900 LBS OF PRESSURE." RECORDS INDICATE THE SUBJECT DEVICE SHIPPED FROM WESTERN TO THE CUSTOMER ON (B)(6) 2011. THE SUBJECT DEVICE WAS NOT RETURNED TO WESTERN FOR EVALUATION, BUT IS BEING RETAINED BY THE CUSTOMER'S ATTORNEYS. WITHOUT PHYSICAL EVALUATION, THE ACTUAL CAUSE OF THE ALLEGED DISCREPANCY CANNOT BE DETERMINED.

Description of Event or Problem · 1

ACCORDING TO INFORMATION PROVIDED BY THE INITIAL REPORTER: OXYGEN REGULATOR GAUGE SHOWED APPROXIMATELY 1000 LBS OF PRESSURE IN OXYGEN TANK BEFORE USE. DURING USE, FLOW CEASED AND ALTERNATE TANK WAS UTILIZED. ASSESSMENT OF OXYGEN TANK AND GAUGE AFTER EVENT REVEALED TANK WAS EMPTY DESPITE REGULATOR GAUGE REFLECTING APPROXIMATELY 800-900 LBS OF PRESSURE. PLAINTIFFS ALLEGE CARDIAC ARREST AND BRAIN INJURY. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
18636 WESTERN MEDICA MEDICAL OXYGEN PRESET REGULATOR CAN WESTERN/SCOTT FETZER CO. M1-540-P

Patients

Seq Age Sex Outcome Treatment
1 20 MO Disability