FDA Adverse Event Malfunction Summary report: N

THE BELMONT RAPID INFUSER, FMS2000

MDR report key: 3750252 · Received January 28, 2014

Report

Report Number
1219702-2014-00001
Event Type
Malfunction
Date Received
January 28, 2014
Date of Event
January 8, 2014
Report Date
January 24, 2014
Manufacturer
BELMONT INSTRUMENT CORP.
Product Code
BSB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED FOR EVAL. UPON RECEIPT, THE FAN GUARDS WERE VERY DIRTY AND THERE WAS A LOT OF BLOOD ON THE HOUSING, SUGGESTING THE UNIT WAS NOT CLEANED AFTER EACH USE AS STATED IN THE OPERATOR'S MANUAL. OTHERWISE, THE UNIT PERFORMED ACCORDING TO OUR SPECIFICATIONS. THE SYSTEM SHUT DOWN DUE TO OVER TEMPERATURE, AS IT IS DESIGNED TO DO: WHENEVER FLUID FROM THE HEAT EXCHANGER REACHES OUR SPECIFIED LIMIT, THE SYSTEM STOPS PUMPING AND HEATING, CLOSES OFF THE LINE TO THE PT AND INSTRUCTS THE USER TO RESTART THE SYSTEM WITH A NEW DISPOSABLE. THE VOLUME OF TUBING BETWEEN THE TEMPERATURE SENSOR AND THE PT IS 45 ML SO THAT NONE OF THE OVER TEMPERATURE FLUID CAN REACH THE PT. THE DISPOSABLE WAS DISCARDED BY THE USER AND THEREFORE NOT RETURNED FOR EVAL.

Description of Event or Problem · 1

ON (B)(6) 2014, IT WAS REPORTED BY THE BIOMED AT THE HOSPITAL, THAT THE BELMONT RAPID INFUSER, FMS2000, EXHIBITED AN "OVER TEMPERATURE" ERROR MESSAGE DURING A LIVER TRANSPLANT. THE SYSTEM WAS REPLACED AND THE CASE WAS CONTINUED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
60243 THE BELMONT RAPID INFUSER, FMS2000 THERMAL INFUSION FLUID WARMER BSB BELMONT INSTRUMENT CORP. FMS2000 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK