FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3741304 · Received April 11, 2014

Report

Report Number
2649622-2014-03126
Event Type
Injury
Date Received
April 11, 2014
Date of Event
December 16, 2013
Report Date
February 3, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY # (B)(4): THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED AND ANALYSIS OF THE DEVICE MEMORY INDICATED PACING CAPTURE THRESHOLD IN THE RV (RIGHT VENTRICLE) WAS ELEVATED. IT WAS NOTED THAT THE ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE TREND ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS DECREASING.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5086MRI52, IMPLANTABLE PACING LEAD, (B)(6) 2013. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAT HIGH THRESHOLDS, A DECREASE IN IMPEDANCE FROM APPROXIMATELY 900-950 OHMS TO APPROXIMATELY 450 OHMS, A POSSIBLE/PROBABLE MICRODISLODGMENT AND DIAPHRAGMATIC STIMULATION. THE RV LEAD WAS REVISED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
225707 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00064 YR Hospitalization| R A2DR01 IMPLANTABLE PACEMAKER