AUTOCAT2 WAVE
Report
- Report Number
- 1219856-2014-00059
- Event Type
- Malfunction
- Date Received
- March 12, 2014
- Date of Event
- March 2, 2014
- Report Date
- March 2, 2014
- Manufacturer
- ARROW INTL., INC.
- Product Code
- DSP
- PMA / PMN Number
- K060309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- NURSE
Narratives
(B)(4). DEVICE WILL NOT BE RETURNED FOR EVAL.
IT WAS REPORTED VIA A CALL ON (B)(6) 2014 THAT THE RN FROM THE INTENSIVE CARE UNIT (ICU) PHONED THE CLINICAL SUPPORT SPECIALIST (CSS) BECAUSE THE PUMP WAS INTERMITTENTLY NOT PUMPING AND GOING FLAT LINED ON BOTH EKG AND ARTERIAL PRESSURE (AP) WHILE PT WAS STABLE WITH PRESSURES AND HRR (HEART RATE REGULAR). THE CSS FIRST CONFIRMED WITH THE RN IN THE ICU THAT THE PT WAS STABLE AND HAD A PULSE. THE RN THEN RELAYED THAT THE PUMP WAS IN AUTOPILOT, PATTERN TRIGGER AND USING THE TRANSDUCER. THE RN STATED IT WAS A FIBEROPTIX SENSOR (FOS) INTRA-AORTIC BALLOON (IAB). THE ICON WAS BLACK AND THE FOS STATUS WAS SHOWING AE. THE CSS CONFIRMED TWICE WITH THE RN THAT THE FOS STATUS WAS AE AND NOT RE. THE CSS ASKED THE RN TO SEND A STRIP FROM THE PUMP SO THE CSS COULD SEE WHAT WAS HAPPENING. UPON RECEIPT OF THE STRIP, IT HAD NO WAVEFORMS PRESENT AND SHOWED NO PRESSURE; HOWEVER, HAD A HR LISTED OF 88 BPM. THE RN HAD THE PUMP SET TO PRINT A STRIP EVERY 15 MINUTES WITH JUST THE AP WAVEFORM. BEST SIGNAL ANALYSIS IS SHOWING A NOISY SIGNAL. THE CSS CALLED THE RN BACK AND ASKED HER TO TRY CHANGING THE PATCHES AND EXPLAINED THAT THE PUMP WAS NOT SEEING THE EKG CLEARLY ALTHOUGH IT MAY LOOK THAT WAY TO HER. THE CSS ALSO SUGGESTED THAT THE RN ATTEMPT TO CHANGE THE EKG CABLE IF THE PATCH CHANGE DIDN'T HELP. THE CSS EXPLAINED TO THE RN AFTER SEEING THE SECOND STRIP THAT THE IAB IS TIMED APPROPRIATELY AND SEEMINGLY SUPPORTING THE PT WELL. THE CSS DID FURTHER DISCUSS THAT THE EKG IS SHOWING NOISE. THE CSS ASKED THE RN TO CHANGE THE PATCHES AND POSSIBLY THE CABLE. THE CSS TOLD THE RN THAT SHE WOULD CALL HER BACK TO FOLLOW UP IN AN HOUR. AT 1754 CST THE CSS CALLED THE RN BACK AND THE RN SAID THAT IT SEEMED TO BE DOING A LITTLE BETTER, BUT WOULD STILL HAVE PAUSES WHERE THE PUMP WOULDN'T PUMP. THE CSS THEN HAD HER DISCONNECT THE EKG CABLE AND THE PUMP IN AUTOPILOT (CONFIRMED BY THE RN) DID NOT SWITCH TO AP AND ALARMED TRIGGER LOSS. THE RN DID CONFIRM THAT THE PT HAD A PRESSURE. THE CSS THEN ASKED THE RN TO MAKE SURE ALL CABLES WERE CONNECTED AND THE PUMP BEGAN PUMPING AGAIN AND DISPLAYED AN AP WAVEFORM. THE CSS ASKED TH ERN IF SHE HAD CHANGED THE EKG CABLE, BUT TH ERN ONLY CHANGED THE PATCHES AS THE CABLE IS ON THE OTHER PUMP, LOCKED IN THE CATH LAB AND THE RN CANNOT GET TO IT. THE CSS THEN ASKED HER TO SAVE THE IAB-05840-LWS ONCE REMOVED AND ALSO SEND THE PUMP TO BIOMED TO BE CHECKED. THE CSS STATED THAT THE PT WAS GOING TO HAVE OPEN HEART SURGERY (OHS) SOMETIME THIS WEEK. THE CSS ASKED THAT THE RN PUT A NOTE ON THE PUMP NAD CATHETER WITH DIRECTIONS AFTER REMOVAL. THE CSS GAVE TH ERN HER DIRECT CELL NUMBER TO GIVE TO NIGHT SHIFT IF THEY SHOULD HAVE ANY QUESTIONS OVERNIGHT. THE CSS RECEIVED NO FURTHER CALLS. ADDITIONAL INFO RECEIVED ON (B)(4) 2014 FROM THE CHARGE NURSE (CN) STATED THAT THE MALE PT WAS TRANSFERRED TO ANOTHER FACILITY ((B)(6)) ON (B)(6) 2014 FOR SURGERY. THE CN SPOKE WITH THE CATH LAB TECHNICIAN WHO FELT IT WAS A FO INTERMITTENT PROBLEM; IT WAS WORKING, BUT NOT CONTINUOUS. THE PUMP WAS NEVER SWITCHED OUT. THE CN ALSO STATED THEY HAD NO ISSUES AFTER THE PHONE CALL TO THE ARROW HOT LINE AND THE FO WAS WORKING APPROPRIATELY WHEN THE PT WAS TRANSFERRED TODAY. THE CN CALLED THE OTHER FACILITY TO SEE IF THE PT HAD GONE INTO SURGERY YET TRYING TO OBTAIN THE IAB AFTER REMOVAL. THE CN CONNECTED ME WITH THE SURGICAL DEPARTMENT. AFTER BEING PASSED AROUND A FEW TIMES I WAS TOLD THEY DID NOT HAVE A PT ON THEIR SCHEDULE THAT WAS TRANSFERRED FROM ANOTHER FACILITY. UNABLE TO RETRIEVE THE RETURN SAMPLE FROM THE SISTER FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149600 | AUTOCAT2 WAVE | INTRA-AORTIC BALLOON PUMP PRODUCTS | DSP | ARROW INTL., INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INTRA-AORTIC BALLOON |