FDA Adverse Event Other Summary report: N

RED LNC-10 PATIENT CABLE

MDR report key: 3736322 · Received March 27, 2014

Report

Report Number
2031172-2014-00028
Event Type
Other
Date Received
March 27, 2014
Date of Event
February 26, 2014
Report Date
February 27, 2014
Manufacturer
MASIMO CORPORATION
Product Code
DSA
PMA / PMN Number
K111888
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE UNIT WAS NOT RETURNED FOR EVALUATION DESPITE MULTIPLE ATTEMPTS. INFO WAS MADE AVAILABLE TO MASIMO INDICATING THAT THE PATIENT IS IN GOOD CONDITION. NO MEDICAL INTERVENTION WAS NECESSARY. THE PATIENT'S PARENTS WERE ABLE TO UNTANGLE THE PATIENT WITHOUT INCIDENT. THE PATIENT CABLE WAS ROUTED OVER THE TOPE OF THE CRIB, NOT THROUGH THE SIDE SLATS. THE DIRECTIONS FOR USE FOR THIS CABLE PROVIDE A WARNING TO "CAREFULLY ROUTE CABLE AND PATIENT CABLE TO REDUCE THE POSSIBILITY OF PATIENT ENTANGLEMENT OR STRANGULATION."

Description of Event or Problem · 1

TWO YEAR OLD CHILD WAS FOUND IN CRIB WITH PATIENT CABLE WRAPPED TWICE AROUND THE PATIENT'S NECK. THE PATIENT MONITOR DID NOT ALARM DURING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
182643 RED LNC-10 PATIENT CABLE DSA MASIMO CORPORATION 2056 13GCM

Patients

Seq Age Sex Outcome Treatment
1 2 YR Other