FDA Adverse Event Malfunction Summary report: N

UNIFY CRT-D, DF-4 CONNECTOR

MDR report key: 3730535 · Received April 7, 2014

Report

Report Number
2938836-2014-08968
Event Type
Malfunction
Date Received
April 7, 2014
Date of Event
March 9, 2012
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. FINAL ANALYSIS CONFIRMED COMPLAINT OF BACKUP OPERATION. THE CAUSE OF THE BACKUP OPERATION WAS DUE TO ENCOUNTERING AN UNCORRECTABLE NMI OR PARITY ERROR IN THE STACK. NO FURTHER ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT THE IMPLANTABLE CARDIAC DEFIBRILLATOR WAS IN BACKUP OPERATION. THE DEVICE WAS NOT IMPLANTED.