FDA Adverse Event Death Summary report: N

PRECISION MEDICAL INC.

MDR report key: 372953 · Received January 16, 2002

Report

Report Number
2523148-2002-00001
Event Type
Death
Date Received
January 16, 2002
Date of Event
January 11, 2002
Report Date
January 16, 2002
Manufacturer
PRECISION MEDICAL, INC.
Product Code
CAX
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

PT WAS ELECTIVELY INTUBATED AFTER THE CONCLUSION OF A CARDIAC CATH PROCEDURE AND RECEIVED NITROUS OXIDE INSTEAD OF OXYGEN. THE OXYGEN FLOWMETER WAS USED WITH AN ADAPTER/CONNECTOR ONE OF TWO PRONGS WAS BROKEN OFF CONNECTOR ALLOWING INSERTION INTO NITROUS OXIDE PORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISION MEDICAL INC. OXYGEN FLOWMETER & CONNECTOR CAX PRECISION MEDICAL, INC. 1MFA4001 NI

Patients

Seq Age Sex Outcome Treatment
1 77 YR Death