FDA Adverse Event
Death
Summary report: N
PRECISION MEDICAL INC.
MDR report key: 372953
·
Received January 16, 2002
Report
- Report Number
- 2523148-2002-00001
- Event Type
- Death
- Date Received
- January 16, 2002
- Date of Event
- January 11, 2002
- Report Date
- January 16, 2002
- Manufacturer
- PRECISION MEDICAL, INC.
- Product Code
- CAX
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
PT WAS ELECTIVELY INTUBATED AFTER THE CONCLUSION OF A CARDIAC CATH PROCEDURE AND RECEIVED NITROUS OXIDE INSTEAD OF OXYGEN. THE OXYGEN FLOWMETER WAS USED WITH AN ADAPTER/CONNECTOR ONE OF TWO PRONGS WAS BROKEN OFF CONNECTOR ALLOWING INSERTION INTO NITROUS OXIDE PORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRECISION MEDICAL INC. | OXYGEN FLOWMETER & CONNECTOR | CAX | PRECISION MEDICAL, INC. | 1MFA4001 | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Death |