FDA Adverse Event Malfunction Summary report: N

THORACIC GRASPER INSTRUMENT

MDR report key: 3729223 · Received April 7, 2014

Report

Report Number
2955842-2014-01998
Event Type
Malfunction
Date Received
April 7, 2014
Date of Event
February 26, 2014
Report Date
March 10, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THE MAIN TUBE INSULATION WAS SCRATCHED AND SOME OF THE INSULATION HAD BEEN REMOVED AT THE DISTAL END. THE SCRATCHES MEASURED APPROXIMATELY .020 - .117 IN LENGTH. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; TUBE ABRASIONS WITH MATERIAL REMOVED, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PULMONARY LOBECTOMY PROCEDURE, BLACK INSULATION ON THE SHAFT OF THE THORACIC GRASPER INSTRUMENT WAS DAMAGED. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
208477 THORACIC GRASPER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420343-01 S10110610 474

Patients

Seq Age Sex Outcome Treatment
1