FDA Adverse Event Malfunction Summary report: N

DRB SSP UNITRAY KIT WITH TAQ POLYMERASE

MDR report key: 3727986 · Received January 23, 2014

Report

Report Number
2244574-2014-00153
Event Type
Malfunction
Date Received
January 23, 2014
Report Date
June 13, 2012
Manufacturer
LIFE TECHNOLOGIES CORPORATION
Product Code
MZI
PMA / PMN Number
BK070038
Removal / Correction Number
004-2012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL INVESTIGATION CONFIRMS CUSTOMER INCIDENT OF LANE 95 BASE PAIR SELECTION NOT LIMITING RESULTS TO EITHER DR4 OR DR5 BASE ON SELECTION OF 85BP OR 210BP RESULT. THIS IS A KNOWN DEFECT OF THE UNIMATCH SOFTWARE WHEN USING A UCH FILE WHERE THE "^" SYMBOLS ARE PRESENT IN THE DRB4/DRB5 MIX. THE PRESENCE OF THE "^" SYMBOL BREAKS THE SIZE FILTERING FUNCTION OF UNIMATCH WHICH RESULTS IN BOTH DRB4 AND DRB5 DISPLAYED IN THE RESULTS REGARDLESS OF THE PRODUCT SIZE ENTERED BY THE USER. DCO1035227 RELEASED 04/06/2012 CONTAINED NEW UCH FILES UPDATING THE REACTIVITY OF DRB1 04:33 IN LANE 44 OF THE DRB1 KIT. THE UCH FILES RELEASED IN DCO1035227 WERE DONE SO WITHOUT THE REQUIRED MANUAL EDITS TO REMOVE THE "^" SYMBOLS IN LANE 95. DISPOSITION IS TO USE AS IS. A CUSTOMER NOTIFICATION WAS SENT ON 05/31/2012 TO AFFECTED CUSTOMERS PROVIDING INSTRUCTIONS ON HOW TO OBTAIN CORRECTED DOCUMENTS FROM THE WEBSITE. ROOT CAUSE IS DETERMINED TO BE AN ERROR IN THE MANUAL PROCESS OF CREATING THIS FUNCTIONALITY WHEN GENERATING THE KIT DATABASE IN HOSII.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING USING THE ALLSET GOLD SSP HLA-DRB1 HIGH RES KIT (CATALOG NUMBER 54040D, LOT NUMBER 017 904113) THAT THE DRB1 LANE 95 BASE PAIR FILTER DID NOT FUNCTION PROPERLY. RESULTS WERE NOT LIMITED TO EITHER DR4 OR DR5. THIS WOULD RESULT IN A MISTYPE. ((B)(4)). THE 4515110 DRB1 SSP UNITRAY KIT WITH TAQ POLYMERASE, LOT NUMBER 017 845070 904691 HAS BEEN IDENTIFIED AS ALSO HAVING THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55093 DRB SSP UNITRAY KIT WITH TAQ POLYMERASE MZI TEST, QUALITATIVE FOR HLA, NON-D MZI LIFE TECHNOLOGIES CORPORATION 017 845070 904691

Patients

Seq Age Sex Outcome Treatment
1