FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3725029 · Received April 3, 2014

Report

Report Number
9611451-2014-00297
Event Type
Malfunction
Date Received
April 3, 2014
Date of Event
March 3, 2014
Report Date
March 7, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE HAVE COMPLETED OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: TWO MR290 AUTOFEED HUMIDIFICATION CHAMBERS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. THE CHAMBERS WERE VISUALLY INSPECTED FOR DAMAGE AND PRESSURE TESTED. RESULTS: NO DAMAGE WAS FOUND TO ANY PART OF THE RETURNED CHAMBERS. THE PRESSURE TEST REVEALED THAT THE CHAMBERS WERE WITHIN SPECIFICATION. UPON THE RETURN OF THE DEVICES THE LOT DATE OF THE MR290 AUTOFEED HUMIDIFICATION CHAMBERS WAS NOTED TO BE 130912 (12 SEPTEMBER 2013). NO COMPLAINTS HAVE BEEN RECEIVED FOR THIS LOT DATE. CONCLUSION: WE COULD FIND NO FAULT WITH THE CHAMBERS AND THEREFORE COULD NOT DETERMINE THE CAUSE OF THE FAULT REPORTED BY THE CUSTOMER. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. CHAMBERS THAT FAIL ANY OF THESE TESTS ARE DISCARDED. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 AUTOFEED HUMIDIFICATION CHAMBERS STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT AN MR290 AUTOFEED HUMIDIFICATION CHAMBER CAUSED AN RT380 ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT TO FAIL THE VENTILATOR LEAK TEST. THIS OCCURRED PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT AN MR290 AUTOFEED HUMIDIFICATION CHAMBER CAUSED AN RT380 ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT TO FAIL THE VENTILATOR LEAK TEST. THIS OCCURRED PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
202919 VENTED AUTOFEED CHAMBER BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED MR290V 130912

Patients

Seq Age Sex Outcome Treatment
1 FPH RT380 ADULT DUAL HEATED EVAQUA2 BREATHING CIRC| FPH RT380 ADULT DUAL HEATED EVAQUA2 BREATHING CIRC